100mm and 200mm Calorimeters Setup and Operating Procedures PN11434C
2Thank you for choosing a Scientech large aperture calorimeter. Scientech, an ISO 9001 registered company, and our employees are pleased to provide
3LARGE APERTURE CALORIMETER SPECIFICATIONS: Model360401 380401 380402 384UV5 360801 380801 380802 384UV5Type of Absorber Surface Volume Volume Volume
4UNPACKING AND SET UP: The calorimeter and accessories are shipped in custom packing materials. All packing materials should be saved for future dam
5Operation of Large Aperture Calorimeters with an Analog Chart Recorder: Calorimeter Response: The response of a calorimeter to a single pulse input
6Initial Voltage Interpolation: A method used to eliminate the tedious numerical integration task is to project the thermal decay envelope on to the
7F. Record the voltage reading of the DVM (Vc). G. Calculate the calorimeter’s output sensitivity (S) as follows: S = Vc/W where: S = calorimeter
8DISPOSAL OF ELECTRICAL AND ELECTRONIC EQUIPMENT: Scientech, Inc. recommends the following for disposal of electrical and electronic equipment: 1. T